Chemyx Nexus Serial Ultra High-Pressure Pumps are designed to solve the toughest fluid delivery problems; from micro-flow to scale-up and pilot plant for chemicals, Oil & Gas, and other industries. The high-pressure pump system provides constant flow, constant pressure, and flow or pressure programming. The system has two configurations: single-pump or dual-pump. With the continuous valve system (CVS), the dual pump system will provide a continuous constant flow operation mode. The piston-driven pump module uses a single-speed gear train. Nitronic 50 is the standard material for cylinders, pistons, and caps, Standard seals are graphite-impregnated Teflon, providing long life under harsh operating conditions.
Technical Specifications
Nexus 75 | Nexus 150 | Nexus 260 | Nexus 350 | Nexus 700 | Nexus 1500 | |
Cylinder Capacity | 2042 | 1000 | 360 | 250 | 127 | 66 |
Power | 100-120 VAC,200-240-VAC ,50/60 Hz | 100-120 VAC,200-240-VAC ,50/60 Hz | 100-120 VAC,200-240-VAC ,50/60 Hz | 100-120 VAC,200-240-VAC ,50/60 Hz | 100-120 VAC,200-240-VAC ,50/60 Hz | 100-120 VAC,200-240-VAC ,50/60 Hz |
Dimensions | 48(D) x 29(W) x 94(H) | 48(D) x 29(W) x 94(H) | 48(D) x 29(W) x 93(H) | 48(D) x 29(W) x 93(H) | 48(D) x 29(W) x 115(H) | 48(D) x 29(W) x 115(H) |
Weight(kg) | 40.8 | 36.5 | 34.0 | 34.9 | 38.0 | 38.1 |
Flow Rate (ml/min) | 0.01-800 | 0.01-400 | 0.001-140 | 0.00001-100 | 0.00001-50 | 0.00001-26 |
Flow Rate Accuracy | ± 0.5% | ± 0.5% | ± 0.3% | ± 0.3% | ± 0.3% | ± 0.3% |
Maximum Seal Leakage | 2 | 1.5 | 1 | 0.5 | 0.25 | 0.2 |
Flow Rate Display Resolution | 0.01 µL/min | 0.01 µL/min | 0.01 µL/min | 0.01 µL/min | 0.01 µL/min | 0.01 µL/min |
Displacement | 74.176 | 36.346 | 13.085 | 9.087 | 4.636 | 2.403 |
Refill Time | 2.5 minutes | 2.5 minutes | 2.5 minutes | 2.5 minutes | 2.5 minutes | 2.5 minutes |
Refill or Depressurization on Rate | 0.01-800 | 0.01-400 | 0.001-140 | 0.00001-100 | 0.00001-50 | 0.00001-26 |
Pressure Range | 0- 1100 psi | 0- 2200 psi | 0- 3800 psi | 0- 5000 psi | 0- 10000 psi | 0- 22000 psi |
Pressure Accuracy | ±0.5 % of full scale at constant temperature | ±0.5 % of full scale at constant temperature | ±0.5 % of full scale at constant temperature | ±0.5 % of full scale at constant temperature | ±0.5 % of full scale at constant temperature | ±0.5 % of full scale at constant temperature |
Pressure Repeatabiility | ±0.5 % of full scale within 48 hours at constant temperature | ±0.5 % of full scale within 48 hours at constant temperature | ±0.5 % of full scale within 48 hours at constant temperature | ±0.5 % of full scale within 48 hours at constant temperature | ±0.5 % of full scale within 48 hours at constant temperature | ±0.5 % of full scale within 48 hours at constant temperature |
Zero Pressure Drift | ±0.25 % of full scale within 48 hours at constant temperature | ±0.25 % of full scale within 48 hours at constant temperature | ±0.25 % of full scale within 48 hours at constant temperature | ±0.25 % of full scale within 48 hours at constant temperature | ±0.25 % of full scale within 48 hours at constant temperature | ±0.25 % of full scale within 48 hours at constant temperature |
Pressure Display Resolution | 0.1 of selected unit | 0.1 of selected unit | 0.1 of selected unit | 0.1 of selected unit | 0.1 of selected unit | 0.1 of selected unit |
Ambient Temperature Range | 5 - 40°C | 5 - 40°C | 5 - 40°C | 5 - 40°C | 5 - 40°C | 5 - 40°C |
Humidity | 95% maximum | 95% maximum | 95% maximum | 95% maximum | 95% maximum | 95% maximum |
Dead(Headspace) Volume(ml) | 5.05 | 2.8 | 1.42 | 1.25 | 1.25 | 1.25 |
Pollution on Degree | 2 | 2 | 2 | 2 | 2 | 2 |
Installation Category | II | II | II | II | II | II |
Maximum Altitude | 2000m | 2000m | 2000m | 2000m | 2000m | 2000m |