Harvard – Minimum And Maximum Flow Rates

Each Chemyx syringe pump system comes with a built-in syringe library that allows you to select from a variety of syringe manufacturers and volumes with predefined syringe barrel diameters. You can also measure your syringe barrel and manually enter it into the pump. The following tables below show the minimum and maximum flow rates for each of our syringe pumps using a Harvard syringe.

Fusion 100 / Fusion 100-X Syringe Pump

Syringe Size I.D. Inner Diameter (mm) Minimum Flow Rate (uL/hr) Maximum Flow Rate (uL/hr)
8mL 9.525 1.28813 803797
20mL 19.13 5.19591 3242250
50mL 28.6 11.6135 7246839

Fusion 200 / Fusion 200-X Syringe Pump

Syringe Size I.D. Inner Diameter (mm) Minimum Flow Rate (uL/hr) Maximum Flow Rate (uL/hr)
8mL 9.525 0.69412 433132
20mL 19.13 2.79986 1747112
50mL 28.6 6.25804 3905018

Fusion 4000 / Fusion 4000-X Syringe Pump

Syringe Size I.D. Inner Diameter (mm) Minimum Flow Rate (uL/hr) Maximum Flow Rate (uL/hr)
8mL 9.525 1.002 761866.3
20mL 19.13 4.0419 3073115
50mL 28.6 9.0341 6868801
100mL 34.9 13.4525 10228212

Fusion 6000 / Fusion 6000-X Syringe Pump

Syringe Size I.D. Inner Diameter (mm) Minimum Flow Rate (uL/hr) Maximum Flow Rate (uL/hr)
8mL 9.525 3.2065 761866.3
20mL 19.13 12.934 3073115
50mL 28.6 28.9091 6868801
100mL 34.9 43.048 10228212

OEM

Syringe Size I.D. Inner Diameter (mm) Minimum Flow Rate (uL/hr) Maximum Flow Rate (uL/hr)
8mL 9.525 2.23433 7268.1538
20mL 19.13 9.01256 29317.30859
50mL 28.6 20.14421 65527.90234