Harvard – Minimum And Maximum Flow Rates

Each Chemyx syringe pump system comes with a built-in syringe library that allows you to select from a variety of syringe manufacturers and volumes with predefined syringe barrel diameters. You can also measure your syringe barrel and manually enter it into the pump. The following tables below show the minimum and maximum flow rates for each of our syringe pumps using a Harvard syringe.

Fusion 100 / Fusion 100-X Syringe Pump

Syringe SizeI.D. Inner Diameter (mm)Minimum Flow Rate (uL/hr)Maximum Flow Rate (uL/hr)
8mL9.5251.28813803797
20mL19.135.195913242250
50mL28.611.61357246839

Fusion 200 / Fusion 200-X Syringe Pump

Syringe SizeI.D. Inner Diameter (mm)Minimum Flow Rate (uL/hr)Maximum Flow Rate (uL/hr)
8mL9.5250.69412433132
20mL19.132.799861747112
50mL28.66.258043905018

Fusion 4000 / Fusion 4000-X Syringe Pump

Syringe SizeI.D. Inner Diameter (mm)Minimum Flow Rate (uL/hr)Maximum Flow Rate (uL/hr)
8mL9.5251.002761866.3
20mL19.134.04193073115
50mL28.69.03416868801
100mL34.913.452510228212

Fusion 6000 / Fusion 6000-X Syringe Pump

Syringe SizeI.D. Inner Diameter (mm)Minimum Flow Rate (uL/hr)Maximum Flow Rate (uL/hr)
8mL9.5253.2065761866.3
20mL19.1312.9343073115
50mL28.628.90916868801
100mL34.943.04810228212

OEM

Syringe SizeI.D. Inner Diameter (mm)Minimum Flow Rate (uL/hr)Maximum Flow Rate (uL/hr)
8mL9.5252.234337268.1538
20mL19.139.0125629317.30859
50mL28.620.1442165527.90234

    Have any questions or comments about this article? Fill out the form below and we'll get back to you as soon as possible!

    Privacy & Terms

    By choosing "I agree" below you agree to Chemyx Inc Terms of Service. You Also agree to our Privacy Policy, which describes how we process your information.